Related Experiment Video
Updated: Apr 4, 2026

Novel and Innovative Hybrid Technique for Type A Aortic Dissection
Published on: March 28, 2025
Comparative Analysis of Sequential Proximal Optimizing Technique Versus Kissing Balloon Inflation Technique in
Gérard Finet1, François Derimay1, Pascal Motreff2
1Department of Interventional Cardiology, Cardiovascular Hospital and Claude Bernard University and INSERM Unités Mixtes de Recherche 1060, Lyon, France.
Objectives:
This study used a fractal bifurcation bench model to compare 6 optimization sequences for coronary bifurcation provisional stenting, including 1 novel sequence without kissing balloon inflation (KBI), comprising initial proximal optimizing technique (POT) + side-branch inflation (SBI) + final POT, called "re-POT."
Background:
In provisional bifurcation stenting, KBI fails to improve the rate of major adverse cardiac events. Proximal geometric deformation increases the rate of in-stent restenosis and target lesion revascularization.
Methods:
A bifurcation bench model was used to compare KBI alone, KBI after POT, KBI with asymmetric inflation pressure after POT, and 2 sequences without KBI: initial POT plus SBI, and initial POT plus SBI with final POT (called "re-POT"). For each protocol, 5 stents were tested using 2 different drug-eluting stent designs: that is, a total of 60 tests.
Results:
Compared with the classic KBI-only sequence and those associating POT with modified KBI, the re-POT sequence gave significantly (p < 0.05) better geometric results: it reduced SB ostium stent-strut obstruction from 23.2 ± 6.0% to 5.6 ± 8.3%, provided perfect proximal stent apposition with almost perfect circularity (ellipticity index reduced from 1.23 ± 0.02 to 1.04 ± 0.01), reduced proximal area overstretch from 24.2 ± 7.6% to 8.0 ± 0.4%, and reduced global strut malapposition from 40 ± 6.2% to 2.6 ± 1.4%.
Conclusions:
In comparison with 5 other techniques, the re-POT sequence significantly optimized the final result of provisional coronary bifurcation stenting, maintaining circular geometry while significantly reducing SB ostium strut obstruction and global strut malapposition. These experimental findings confirm that provisional stenting may be optimized more effectively without KBI using re-POT.
More Related Videos
13:48Reduction of Radiation Exposure during Endovascular Treatment of Peripheral Arterial Disease Combining Fiber Optic RealShape Technology and Intravascular Ultrasound
Published on: April 21, 2023
12:00Evaluation of a Novel Laser-assisted Coronary Anastomotic Connector - the Trinity Clip - in a Porcine Off-pump Bypass Model