Comparison of two devices for measuring exercise transcutaneous oxygen pressures in patients with claudication
Marie Urban1, Alban Fouasson-Chailloux, Isabelle Signolet
11 University Hospital, Angers, France.
Background:
We aimed at estimating the agreement between the Medicap" (photo-optical) and Radiometer• (electrochemical) sensors during exercise transcutaneous oxygen pressure ( tcp02) tests. Our hypothesis was that although absolute starting values (tcp02rest: mean over 2 minutes) might)e different, tcpOTchanges over time and the minimal value of the decrease from rest of oxygen pressure (DROPmin) results at exercise shall be concordant between the two systems. taneously, one of each system on the chest, on each buttock and on each calf.
Results:
Seventeen Medicap" probes disconnected during the tests. tcp02rest and DROPmin values were higher with Medicap• than with Radiometer•, by 13.7 ± 17.1mmHg and 3.4 ± 11.7 mmHg, respectively. Despite the differences in absolute starting values, changes over time were similar between the two systems. The concordance between the two systems was approximately 70% for classification of test results from DROPmin.
Conclusions:
Photo-optical sensors are promising alternatives to electro-chemical sensors for exercise oximetry, provided that miniaturization and weight reduction of the new sensors are possible.
Patients And Methods:
Forty seven patients with arterial claudication ( 65 ± 7 years) performed a treadmill test with 5 probes each of the electro-chemical and photo-optical devices simul-
More Related Videos
Related Concept Videos
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Peripheral Arterial Disease II: Clinical Manifestations and Diagnostic Evaluation
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Assessing Blood pressure in the Leg
Preparation: