Motion artifacts in kidney stone imaging using single-source and dual-source dual-energy CT scanners: a phantom study
El-Sayed H Ibrahim1,2, Joseph G Cernigliaro3, Robert A Pooley3
1Mayo Clinic, 4500 San Pablo Rd, Jacksonville, FL, 32224, USA. elsayei@umich.edu.
Motion artifacts significantly impact dual-energy computed tomography (DECT) stone classification. Dual-source DECT (DS-DECT) is more robust against motion artifacts than single-source DECT (SS-DECT), reducing misclassification rates.
Area of Science:
- Medical Imaging
- Radiology
- Nephrology
Background:
- Dual-energy computed tomography (DECT) accurately differentiates uric acid from non-uric acid kidney stones.
- Dual-source (DS) scanners acquire low- and high-energy images simultaneously, while single-source (SS) scanners acquire them sequentially.
- Motion artifacts can compromise image quality and diagnostic accuracy in DECT.
Purpose of the Study:
- To evaluate the impact of motion artifacts on kidney stone classification using both SS-DECT and DS-DECT.
- To compare the accuracy of stone classification between SS-DECT and DS-DECT under induced motion conditions.
Main Methods:
- 114 kidney stones were imaged using both DS-DECT and SS-DECT (80 and 140 kVp) with and without induced motion.
- Material-specific images were generated through postprocessing.
- Dual-energy ratio (DER) and stone material were determined and compared across scan types.
Main Results:
- Without motion, SS-DECT correctly classified all stones, while DS-DECT misclassified two cystine stones.
- With induced motion, SS-DECT misclassified 94% of stones, compared to 11% with DS-DECT (P < 0.0001).
- Stone size did not influence misclassification; stone type affected misclassification more with DS-DECT (cystine).
Conclusions:
- Motion artifacts can lead to kidney stone misclassification in DECT.
- DS-DECT demonstrates superior resilience to motion artifacts compared to SS-DECT.
- Proximity of stones and potential miscounting of stones in DS-DECT are significant considerations.
Related Concept Videos
Imaging Studies III: Computed Tomography
Imaging Studies I: Kidney, Ureter, and Bladder Studies
Imaging Studies II: Ultrasonography
Imaging Studies IV: Magnetic Resonance Imaging
Imaging Studies for Cardiovascular System V: CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...