Evaluation of an empirical monitor output estimation in carbon ion radiotherapy
Akihiko Matsumura1, Ken Yusa1, Tatsuaki Kanai1
1Gunma University Heavy Ion Medical Center, Gunma University, 3-39-22 Showa-machi, Maebashi, Gunma 371-8511, Japan.
Purpose:
A conventional broad beam method is applied to carbon ion radiotherapy at Gunma University Heavy Ion Medical Center. According to this method, accelerated carbon ions are scattered by various beam line devices to form 3D dose distribution. The physical dose per monitor unit (d/MU) at the isocenter, therefore, depends on beam line parameters and should be calibrated by a measurement in clinical practice. This study aims to develop a calculation algorithm for d/MU using beam line parameters.
Methods:
Two major factors, the range shifter dependence and the field aperture effect, are measured via PinPoint chamber in a water phantom, which is an identical setup as that used for monitor calibration in clinical practice. An empirical monitor calibration method based on measurement results is developed using a simple algorithm utilizing a linear function and a double Gaussian pencil beam distribution to express the range shifter dependence and the field aperture effect.
Results:
The range shifter dependence and the field aperture effect are evaluated to have errors of 0.2% and 0.5%, respectively. The proposed method has successfully estimated d/MU with a difference of less than 1% with respect to the measurement results. Taking the measurement deviation of about 0.3% into account, this result is sufficiently accurate for clinical applications.
Conclusions:
An empirical procedure to estimate d/MU with a simple algorithm is established in this research. This procedure allows them to use the beam time for more treatments, quality assurances, and other research endeavors.
More Related Videos
07:31Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
10:33Expedited Radiation Biodosimetry by Automated Dicentric Chromosome Identification ADCI and Dose Estimation
Published on: September 4, 2017
Related Concept Videos
Biological Effects of Radiation
Radiation: Applications
The average...