Failure of communication and capture: The perils of temporary unipolar pacing system
Efe Sahinoglu1, Thomas J Wool2, Kenneth J Wool3
1University of Alabama (UAB) School of Medicine, VH 227, 1530 3rd Ave S Birmingham, AL 35294, United States.
Abstract:
We present a case of a patient with pacemaker dependence secondary to complete heart block who developed loss of capture of her temporary pacemaker. Patient developed torsades de pointes then ventricular fibrillation, requiring CPR and external cardioversion. After patient was stabilized, it was noticed that loss of capture of pacemaker corresponded with nursing care, when the pulse generator was lifted off patient׳s chest wall, and that patient׳s temporary pacing system had been programmed to unipolar mode without knowledge of attending cardiologist. This case highlights the importance of communication ensuring all caregivers are aware of mode of the temporary pacing system.