Related Experiment Video
Updated: Apr 4, 2026

Author Spotlight: Microbial Control and Monitoring Strategies for Cleanroom Environments and Cellular Therapies
Published on: March 17, 2023
Sterility maintenance study: Dynamic evaluation of sterilized rigid containers and wrapped instrument trays to
Harry L Shaffer1, Delbert A Harnish2, Michael McDonald2
1Sterilization Consulting Services, LLC, Highlands Ranch, CO.
Background:
Sterilized packaging systems are designed to maintain the sterility of surgical instruments and devices from the time of sterilization until use. This study evaluated the effectiveness of rigid containers versus wrapped instrument trays, sterilized using North American sterilization protocols, to maintain a sterile internal environment poststerilization when challenged with aerosolized bacteria under dynamic environmental conditions.
Methods:
Using a custom aerosol chamber, 111 rigid containers of various durations of use (unused, used <5 years, used 5-9 years) and 161 wrapped trays using 3 grades of sterilization wrap were challenged with ~10(2) colony-forming units per liter of air containing aerosolized Micrococcus luteus with a count median particle size of 1 μm, while simultaneously experiencing air volume exchanges due to vacuum cycles-two 1-psi cycles, three 0.7-psi cycles, and three 0.4-psi cycles-to simulate air exchange events occurring during the sterilization, transportation, and storage of sterilized instrument trays in health care facilities.
Results:
Of 111 rigid containers tested, 97 (87%) demonstrated bacterial ingress into the container. Of 161 wrapped trays, 0 (0%) demonstrated bacterial ingress into the tray. Contamination rates of rigid containers increased significantly with increasing duration of use.
Conclusions:
In this study using a dynamic bacterial aerosol challenge, sterilized wrapped trays demonstrated significantly greater protection than sterilized rigid containers against the ingress of airborne bacteria.
Related Concept Videos
Key Techniques in Microbiology
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Scale-Up Processes