A Method to Improve Electron Density Measurement of Cone-Beam CT Using Dual Energy Technique
Kuo Men1, Jian-Rong Dai1, Ming-Hui Li1
1Department of Radiation Oncology, Cancer Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing 100021, China.
Purpose:
To develop a dual energy imaging method to improve the accuracy of electron density measurement with a cone-beam CT (CBCT) device.
Materials And Methods:
The imaging system is the XVI CBCT system on Elekta Synergy linac. Projection data were acquired with the high and low energy X-ray, respectively, to set up a basis material decomposition model. Virtual phantom simulation and phantoms experiments were carried out for quantitative evaluation of the method. Phantoms were also scanned twice with the high and low energy X-ray, respectively. The data were decomposed into projections of the two basis material coefficients according to the model set up earlier. The two sets of decomposed projections were used to reconstruct CBCT images of the basis material coefficients. Then, the images of electron densities were calculated with these CBCT images.
Results:
The difference between the calculated and theoretical values was within 2% and the correlation coefficient of them was about 1.0. The dual energy imaging method obtained more accurate electron density values and reduced the beam hardening artifacts obviously.
Conclusion:
A novel dual energy CBCT imaging method to calculate the electron densities was developed. It can acquire more accurate values and provide a platform potentially for dose calculation.
More Related Videos
10:42In Depth Analyses of LEDs by a Combination of X-ray Computed Tomography CT and Light Microscopy LM Correlated with Scanning Electron Microscopy SEM
Published on: June 16, 2016
06:56Author Spotlight: Advancements in X-ray CT Tool Chain for Tree Core Analysis
Published on: September 22, 2023
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Electron Microscope Tomography and Single-particle Reconstruction
Electron Tomography
Electron tomography can be performed either in TEM or STEM (scanning transmission...
Scanning Electron Microscopy
Fundamental Principles
Accelerated...