Related Experiment Video
Updated: Apr 4, 2026

A New Murine Model of Endovascular Aortic Aneurysm Repair
Published on: July 7, 2013
Deformable Surface Model for the Evaluation of Abdominal Aortic Aneurysms Treated with an Endovascular Sealing System
Mariano E Casciaro1,2, Salma El-Batti3,4,5, Gilles Chironi6,7,4
1Facultad de Ingeniería Ciencias Exactas y Naturales, Favaloro University, Solís 453, 1078, Ciudad Autónoma de Buenos Aires, Argentina.
Abstract:
Rupture of abdominal aortic aneurysms (AAA) is responsible for 1-3% of all deaths among the elderly population in developed countries. A novel endograft proposes an endovascular aneurysm sealing (EVAS) system that isolates the aneurysm wall from blood flow using a polymer-filled endobag that surrounds two balloon-expandable stents. The volume of injected polymer is determined by monitoring the endobag pressure but the final AAA expansion remains unknown. We conceived and developed a fully deformable surface model for the comparison of pre-operative sac lumen size and final endobag size (measured using a follow-up scan) with the volume of injected polymer. Computed tomography images were acquired for eight patients. Aneurysms were manually and automatically segmented twice by the same observer. The injected polymer volume resulted 9% higher than the aneurysm pre-operative lumen size (p < 0.05), and 11% lower than the final follow-up endobag volume (p < 0.01). The automated method required minimal user interaction; it was fast and used a single set of parameters for all subjects. Intra-observer and manual vs. automated variability of measured volumes were 0.35 ± 2.11 and 0.07 ± 3.04 mL, respectively. Deformable surface models were used to quantify AAA size and showed that EVAS system devices tended to expand the sac lumen size.
More Related Videos
04:56Modified Octopus Technique for Thoracoabdominal Aortic Aneurysm
Published on: August 1, 2025
13:48Reduction of Radiation Exposure during Endovascular Treatment of Peripheral Arterial Disease Combining Fiber Optic RealShape Technology and Intravascular Ultrasound
Published on: April 21, 2023