Related Experiment Video
Updated: Apr 4, 2026

Development and Standardization of an Ex Vivo Micromethod for Intracellular Quantification of Vincristine in Primary ALL Cells by LC-MS/MS
Published on: January 23, 2026
Determination of Voriconazole Concentrations in Serum by GC-MS
Alisha Smith1, Van Leung-Pineda2
1Department of Pathology and Laboratory, Cook Children's Medical Center, Fort Worth, Texas.
Background:
Voriconazole is a broad spectrum triazole antifungal drug used to treat systemic fungal infections. Therapeutic drug monitoring of voriconazole is necessary for achieving maximal efficiency without inducing toxic side effects. Other publications have reported methods for measuring voriconazole in serum using high-performance liquid chromatography (HPLC) and liquid chromatography tandem mass spectrometry (LC-MS/MS). Here, we report for the first time a method for the measurement of voriconazole in serum samples using gas chromatography mass spectrometry (GC-MS).
Methods:
Protein precipitation with methanol was used to extract the antifungal that was derivatized with BSTFA (Sigma-Aldrich, St. Louis, MO) and analyzed by GC-MS. Linearity, sensitivity, precision, accuracy, and drug interferences were evaluated for this assay.
Results:
Our method was linear up to 10 μg/ml of voriconazole. The LOQ was determined to be 0.4 μg/ml. CV for between-day precision was <12%. Correlation with an established LC-MS/MS yielded a R2 of 0.96. Tested drugs did not result in >10% error in measurement.
Conclusion:
To our knowledge, we report here the first GC-MS method for voriconazole measurement with acceptable performance. We hope that this method allows clinical laboratories without HPLC or LC-MS/MS instrumentation to measure voriconazole.
More Related Videos
Related Concept Videos
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall....
Determination of Multiple Dosing Parameters: Steady-State, Minimum and Maximum Concentrations
Drug Concentrations: Measurements
Plasma...
Drug Concentration Versus Time Correlation
Two pivotal parameters are the minimum effective concentration (MEC) and the minimum toxic concentration (MTC). The MEC is the...
Therapeutic Drug Monitoring: Drug Analysis Methods