Related Experiment Video
Updated: Apr 4, 2026

4D Printed Bifurcated Stents with Kirigami-Inspired Structures
Published on: July 25, 2019
Geometrical characteristics after Y-stenting of the basilar bifurcation
Muzaffer Sağlam1, Osman Kızılkılıç, Vania Anagnostakou
1Department of Radiology, GATA Haydarpaşa Training and Research Hospital, İstanbul, Turkey. mzsaglam@yahoo.com.
Purpose:
We aimed to investigate the angular changes after Y-stenting of the basilar bifurcation aneurysms.
Methods:
A total of 19 patients (age range, 27-80 years; mean age, 52.5 years) underwent Y-stent coiling for basilar bifurcation aneurysm. Three vascular angles (α, β1, β2) were measured in the anteroposterior plane. β1 and β2 represented the angles between the basilar artery and the proximal P1 segments of the right and left posterior cerebral arteries, respectively. α represented the complementary angle between the β1 and β2 angles. Angles were measured before and after stent deployment. Diameters of the basilar artery and P2 segment of the posterior cerebral artery were measured at both sides. Correlation between vascular diameter and angular change of the basilar bifurcation was investigated.
Results:
Statistically significant α, β1, and β2 angle changes were found after stent deployment (P < 0.001). There was no statistically significant relationship between the diameter of the basilar artery and the α, β1, β2 angle changes (P > 0.05). There was no statistically significant relationship between the diameter of the posterior cerebral artery and the β angle change (P > 0.05). We found a statistically significant inverse correlation between pre-stent β angle and post-stent angle change (right side, P = 0.008; left side, P < 0.001).
Conclusion:
Y-stenting narrows the effective neck and straightens the vascular bifurcation angle. Most of the angular remodeling occurs on the side that had a more acute angle before stent deployment.
More Related Videos
07:04Creation of Two Saccular Elastase-Digested Aneurysms with Different Hemodynamics in One Rabbit
Published on: April 15, 2021
11:00Experimental Investigation of Secondary Flow Structures Downstream of a Model Type IV Stent Failure in a 180° Curved Artery Test Section
Published on: July 19, 2016
Related Concept Videos
Mitral Stenosis I: Introduction
Deformations in a Symmetric Member in Bending
When the member is segmented into tiny cubic elements, it is observed that the primary stress...
Mitral Stenosis II: Clinical features and Diagnostic Tests
Bending of Curved Members - Strain Analysis
The important part of bending analysis for such a member...