Related Experiment Video
Updated: Apr 4, 2026

Evaluation of a Novel Laser-assisted Coronary Anastomotic Connector - the Trinity Clip - in a Porcine Off-pump Bypass Model
Published on: November 24, 2014
Patent ductus arteriosus occlusion in small dogs utilizing a low profile Amplatz® canine duct occluder prototype
Christopher D Stauthammer1, Janet Olson1, Damon Leeder1
1Department of Veterinary Clinical Sciences, College of Veterinary Medicine, University of Minnesota, St. Paul, MN, USA.
Objectives:
To develop procedural methodology and assess the safety, utility and effectiveness of a low profile Amplatz(®) canine duct occluder (ACDO) prototype in dogs deemed too small to undergo ductal occlusion with the commercially-available ACDO device.
Animals:
Twenty-one dogs with left-to-right shunting patent ductus arteriosus (PDA). Dogs were ≥1.5 kg but considered too small to accommodate a 6 Fr catheter or 4 Fr sheath within the femoral artery.
Methods:
Prospective canine study using a low profile ACDO prototype delivered through a 4 Fr catheter via a femoral arterial approach. Procedural methods, fluoroscopy time, perioperative complications, and residual ductal flow were evaluated, and angiographic ductal morphology and dimensions were tabulated.
Results:
All 21 dogs underwent successful ductal occlusion using the prototype device, 4 Fr catheter, and right femoral artery approach. No perioperative complications or device embolization occurred. The median minimal ductal diameter was 1.9 mm (range, 0.4-3.4), and the median device size deployed was 4 mm (range, 3-6). Complete ductal occlusion was noted in 17 dogs (81%) on post-deployment angiography. Twenty dogs (95%) had no residual flow on echocardiography performed the following day. In the 17 dogs (81%) that returned for a long-term (≥3months) follow-up evaluation, all had complete ductal occlusion based on echocardiography.
Conclusions:
The low profile ACDO prototype is a safe and effective method of PDA occlusion in the small dog. The deployment procedure appears of similar technical difficulty to the commercially available ACDO.