A rapid method for measuring intraocular lens power in vitro with a focimeter
Mari Carmen García-Domene1, María Amparo Díez-Ajenjo1, Cristina Peris-Martínez2
1FISABIO-Oftalmología Médica, Bifurcación Pío Baroja-General Aviles, s/n, 46015 Valencia, Spain; Universitat de València, C/Dr. Moliner, 50, 46100 Burjassot, Valencia, Spain.
Experimental Eye Research
|September 20, 2015
Abstract:
In this paper we describe a new method for measuring the intraocular lens (IOL) power using a focimeter, a negative ophthalmic lens and a saline solution (0.9% NaCl). To test this we measured the power of 58 different IOLs and we compared them with the power stated by the manufacturer. Despite the limitations, the results show a good correlation.