Development of a New Imaging Bulk Package for Multipatient Use in MDCT
Alberto Spinazzi1, Patrice A Marchildon2, Elyssa A Murray2
11 Global Medical and Regulatory Affairs, Bracco Diagnostics, Inc., Monroe Township, NJ.
Objective:
Multidose presentations of U.S. Food and Drug Administration (FDA)-approved radiographic contrast agents have been considered pharmacy bulk packages. However, the use of pharmacy bulk packages for multipatient dosing does not meet the U.S. Pharmacopeia definition of a pharmacy bulk package. The purpose of this study was to validate and gain FDA approval for a new multidose preparation of iopamidol for safe, compliant multipatient dosing in the CT suite.
Materials And Methods:
An FDA-approved development program was undertaken to determine whether multidose presentations of iopamidol used in combination with a transfer set remain free of chemical and microbiologic contamination during the labeled maximum hold time after container closure penetration and simulated worst-case handling conditions. The program comprised antimicrobial effectiveness testing of iopamidol-300 and iopamidol-370 containers with seven microbes. Microbial growth was evaluated at five time points up to 28 days after introduction. Microbial ingress testing involved inoculation of four challenge sites with each of four microorganisms for up to 14 hours. Chemical compatibility and extractable testing was performed to ensure chemical integrity.
Results:
No growth of microorganisms occurred. All evaluated samples remained sterile, indicating no microbial contamination through 14 hours of simulated clinical use. No effect on chemical integrity was found in any of the drawn iopamidol samples meeting the chemical specifications for iopamidol, and no leachable compounds were detected.
Conclusion:
The absence of any chemical or microbiologic contamination led the FDA to approve the iopamidol multidose container and transfer set as a combination product for multipatient use. The approval resulted in a new U.S. Pharmacopeia category of multidose presentation-the imaging bulk package.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System V: CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...