Reliability tests and guidelines for B-mode ultrasound assessment of central adiposity
Lee Stoner1, Victoria Chinn2, Jon Cornwall3
1School of Sport and Exercise, Massey University, Wellington, New Zealand.
Introduction:
Ultrasound represents a validated and relatively inexpensive diagnostic device for assessing central adiposity; however, widespread adoption has been impeded by the lack of reliable standard operating procedures.
Purpose:
To examine the reliability of, and describe guidelines for, ultrasound-derived recording of intra-abdominal fat thickness (IAT) and maximal preperitoneal fat thickness (PFT).
Methods:
Ultrasound scans were obtained from 20 adults (50% female, 26 ± 7 years, 24·5 kg/m(2) ) on three different mornings. IAT was assessed 2 cm above the umbilicus (transverse plane) measuring from linea alba to: (i) anterior aorta, (ii) posterior aorta and (iii) anterior aspect of the vertebral column. PFT was measured from linea alba to visceral peritoneum in (i) sagittal and (ii) transverse planes, immediately over and inferior to the xiphi-sternum, respectively.
Results:
For IAT, the criterion intraclass correlation coefficient (ICC) of 0·75 was exceeded for measurements to anterior aorta (0·95), posterior aorta (0·94) and vertebra (0·96). The reliability coefficient expressed as a percentage of the mean (RC%) was lowest (better) for measurement to vertebrae (9·8%). For PFT, mean thickness was comparable for sagittal (1·74 cm) and transverse (1·76 cm) planes; ICC values were also comparable for both planes (0·98 vs. 0·98, respectively), as were RC% (7·5% vs. 7·1%, respectively).
Conclusions:
IAT assessments to the vertebra were marginally more reliable than those to other structures. While PFT assessments were equally reliable for both measurements planes, precise probe placement was easier for the sagittal plane. Based on these findings, guidelines for the reliable measurement of central adiposity using ultrasound are presented.
Related Concept Videos
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Ultrasound I: Abdominal Ultrasonography
Abdominal ultrasonography, commonly known as abdominal ultrasound, is a vital, non-invasive medical imaging technique widely used in healthcare.
Procedure:
This diagnostic tool allows the clinician to visually inspect internal structures within the abdomen, including vital organs such as the liver, gallbladder, pancreas, kidneys, and spleen.
The abdominal ultrasound process begins with applying a special gel to the patient's skin over the abdomen. This gel enhances the...
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):
Ultrasonography
During an ultrasonography procedure, a handheld device called...