Stability indicating RP-HPLC method for simultaneous determination of piroxicam and ofloxacin in binary combination
Peter John1, Waqar Azeem1, Muhammad Ashfaq2
1Department of Chemistry, Government College University, Lahore, Pakistan.
Abstract:
A simple and precise RP-HPLC method was developed for simultaneous determination of piroxicam and ofloxacin in pharmaceutical formulations and human serum. Optimum separations of piroxicam, ofloxacin and stress-induced degradation products were achieved by use of Hypersil BDS C8 column (250 x 4.6mm, 5 μm). The mobile phase was a mixture of acetonitrile: 0.012M K2HPO4: 0.008M sodium citrate (both buffers mixed and pH adjusted to 2.8) (50:25:25 v/v/v) delivered at flow rate of 1.5 mL min⁻¹ using DAD at 254 nm. Response was linear function of concentration over the ranges of 70-130 mg mL⁻¹ for piroxicam and ofloxacin (r² ≥ 0.999). The method efficiently separated the analytical peaks from degradation products with acceptable tailing and resolution. The developed method was successfully used for concurrent analysis of piroxicam and ofloxacin in pharmaceutical formulations, human serum and in vitro drug interaction studies.
More Related Videos
08:57Sample Extraction and Simultaneous Chromatographic Quantitation of Doxorubicin and Mitomycin C Following Drug Combination Delivery in Nanoparticles to Tumor-bearing Mice
Published on: October 5, 2017
11:17Multiplex Therapeutic Drug Monitoring by Isotope-dilution HPLC-MS/MS of Antibiotics in Critical Illnesses
Published on: August 30, 2018