Dosimetric characterization and image quality evaluation of the AIRO mobile CT scanner
Victor J Weir1, Jie Zhang2, Angela P Bruner2
1Baylor Health Care System, Suite Dallas, TX, USA.
Abstract:
Radiation dose and image quality from a recently introduced mobile CT imaging system are presented. Radiation dose was measured using a conventional 100 mm pencil ionization chamber and CT polymethylmetacrylate (PMMA) body and head phantoms. Image quality was evaluated with a CATPHAN 500 phantom. Spatial resolution, low contrast resolution, Modulation Transfer Function (MTF), and Normalized Noise Power Spectrum (NNPS) were analyzed. Radiation dose and image quality were compared to those from a multi-detector CT scanner (Siemens Sensation 64). Under identical technique factors radiation dose (mGy/mAs) from the AIRO mobile CT system (AIRO) is higher than that from a 64 slice CT scanner. Based on MTF analysis, both Soft and Standard filters of the AIRO system lost resolution quickly compared to the Sensation 64 slice CT. The Siemens scanner had up to 7 lp/cm for the head FOV and H40 kernel and up to 5 lp/cm at body FOV for the B40f kernel. The Standard kernel in the AIRO system was evaluated to have 3 lp/cm and 4 lp/cm for the body and head FOVs respectively. NNPS of the AIRO shows low frequency noise due to ring-like artifacts which may be caused by detector calibration or lack of artifact reducing image post-processing. Due to a higher dose in terms of mGy/mAs at both head and body FOV, the contrast to noise ratio is higher in the AIRO system than in the Siemens scanner. However detectability of the low contrast objects is poorer in the AIRO due to the presence of ring artifacts in the location of the targets.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System V: CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Radiological Investigation I: X-ray and CT
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...