Evaluation of varying physical acquisition parameters in gamma camera gated cardiac SPECT
Reem H Elgebaly1, Monira M Rageh1, Mahmoud Adel2
1Department of Biophysics, Faculty of science, Cairo University, Cairo, Giza, Egypt.
Background:
There are various physical factors that limit the quality of the Single Photon Emission Computed Tomography (SPECT) images such as number of frames per projection, number of projections per scan and time per projection in gated myocardial SPECT scan.
Objective:
The aim of the present work was to evaluate physical acquisition parameters affecting the cardiac imaging to optimize the patient's examination time and image quality that are acquired with the gamma camera and workers radiation protection (through reducing time of radiation exposure) in cardiac SPECT facility. Cardiac functional parameters like ejection fraction (EF), end diastole volume (EDV), end systole volume (ESV) and total perfusion defect (TPD) were evaluated.
Methods:
Thirty patients were divided into three groups as follows: 1- Group (1): contains two sub-groups, one scanned with 8 frames per projection, the other with 16 frames per projection. 2- Group (2): contains two sub-groups, one scanned with 32 projections per scan, the other with 16 projections per scan. 3- Group (3): contains two sub-groups, one scanned at a time per projection of 20 s and the other scanned at 10 s per projection. Parameters such as the Ejection Fraction (EF), End Diastole Volume (EDV), End Systole Volume (ESV) and Total Perfusion Defect (TPD) in SPECT were compared for each group.
Results:
There was insignificant difference in the EF, EDV, ESV and TPD between the different groups concerning the number of frames per projection, number of projections per scan and time per projection.
Conclusion:
The reduction of scan time to half or the number of projections per scan to half or increasing the number of frames per scan from 8 to 16 at the same duration has insignificant effect in the cardiac functional parameters used with gated SPECT scan.
More Related Videos
Related Concept Videos
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion,...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...