Dual Energy CT Angiography of Peripheral Arterial Disease: Feasibility of Using Lower Contrast Medium Volume
Abdulrahman Almutairi1, Zhonghua Sun2, Abduljaleel Poovathumkadavi3
1Department of Medical Radiation Sciences, School of Science, Curtin University, Perth, 6845, Western Australia, Australia; Department of Medical Imaging, King Fahad Specialist Hospital, Dammam, 31444, Saudi Arabia.
Objective:
One of the main drawbacks associated with Dual Energy Computed Tomography Angiography (DECTA) is the risk of developing contrast medium-induced nephropathy (CIN). The aim of the present study was firstly, to design an optimal CT imaging protocol by determining the feasibility of using a reduced contrast medium volume in peripheral arterial DECTA, and secondly, to compare the results with those obtained from using routine contrast medium volume.
Methods:
Thirty four patients underwent DECTA for the diagnosis of peripheral arterial disease. They were randomly divided into two groups: Group 1 (routine contrast volume group) with n = 17, injection rate 4-5 ml/s, and 1.5 ml/kg of contrast medium, and Group 2 ((low contrast volume group), with n = 17, injection rate 4-5ml/s, and contrast medium volume 0.75 ml/kg. A fast kilovoltage-switching 64-slice CT scanner in the dual-energy mode was employed for the study. A total of 6 datasets of monochromatic images at 50, 55, 60, 65, 70 and 75 keV levels were reconstructed with adaptive statistical iterative reconstruction (ASIR) at 50%. A 4-point scale was the tool for qualitative analysis of results. The two groups were compared and assessed quantitatively for image quality on the basis of signal-to-noise ratio (SNR) and contrast-to-noise-ratio (CNR). Radiation and contrast medium doses were also compared.
Results:
The overall mean CT attenuation and mean noise for all lower extremity body parts was significantly lower for the low volume contrast group (p<0.001), and varied significantly between groups (p = 0.001), body parts (p<0.001) and keVs (p<0.001). The interaction between group body parts was significant with CT attenuation and CNR (p = 0.002 and 0.003 respectively), and marginally significant with SNR (p = 0.047), with minimal changes noticed between the two groups. Group 2 (low contrast volume group) displayed the lowest image noise between 65 and 70 keV, recorded the highest SNR and CNR at 65 keV, and produced significantly lower results with respect to contrast medium volume and duration of contrast injection (p<0.001). The effect of radiation dose was not statistically significant between the two groups.
Conclusions:
DECTA images created at 65 keV and 50% ASIR with low contrast medium volume protocol, yielded results that were comparable to routine contrast medium volume, with acceptable diagnostic images produced during the evaluation of peripheral arteries.
More Related Videos
13:48Reduction of Radiation Exposure during Endovascular Treatment of Peripheral Arterial Disease Combining Fiber Optic RealShape Technology and Intravascular Ultrasound
Published on: April 21, 2023
04:35Sample Preparation for Computed Tomography-based Three-dimensional Visualization of Murine Hind-limb Vessels
Published on: October 7, 2021
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Peripheral Arterial Disease II: Clinical Manifestations and Diagnostic Evaluation
Radiological Investigation I: X-ray and CT
Imaging Studies III: Computed Tomography
Imaging Studies VII: Vascular Imaging
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...