Dual-Energy Computed Tomography-How Accurate Is Gemstone Spectrum Imaging Metal Artefact Reduction? Its Application
Danè Dabirrahmani1, John Magnussen, Richard Charles Appleyard
1From the Australian School of Advanced Medicine, Macquarie University, Sydney, Australia.
Objective:
To assess the accuracy and suitability of dual-energy computed tomography (DECT) in scanning metals used in orthopedic implants.
Materials And Methods:
Four metal phantoms (Cobalt Chrome, Titanium Grade 5, Stainless Steel 316, and Stainless Steel 630), commonly used materials in orthopedic implants, were scanned by conventional, polychromatic CT as well as Gemstone Spectrum Imaging (GSI) DECT, with and without metal artefact reduction software (MARS). Scans were assessed for artefact based on Hounsfield unit values; and surfaces generated, based on a Canny edge detection algorithm. Two separate metal implants were also scanned and assessed for dimensional accuracy.
Results:
Conventional, polychromatic CT, and GSI DECT (without MARS) scans displayed major beam hardening in the presence of all four metals. The GSI DECT with MARS showed very clear and reproducible boundaries with minimal noise surrounding the metal phantoms. However, geometric analysis found overestimation of the dimensions, volume, and surface area for most of the metal phantoms. Titanium displayed the least artefact, compared to the other metals, in all scan scenarios.
Conclusions:
Although metal artefact reduction using GSI DECT looks superior to conventional CT, when measured objectively, it was shown to overestimate geometries and skew dimensions. The GSI DECT with MARS should be used with caution, especially when assessing questions of implant shape or wear.
More Related Videos
07:48High Spatial Resolution Chemical Imaging of Implant-Associated Infections with X-ray Excited Luminescence Chemical Imaging Through Tissue
Published on: September 30, 2022
09:32Time-Resolved, Dynamic Computed Tomography Angiography for Characterization of Aortic Endoleaks and Treatment Guidance via 2D-3D Fusion-Imaging
Published on: December 9, 2021
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
X-ray Imaging
Imaging Studies III: Computed Tomography
Radiological Investigation I: X-ray and CT
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...