Evaluation of Capnography Sampling Line Compatibility and Accuracy when Used with a Portable Capnography Monitor
Published on: September 29, 2020
Diagnostic accuracy of capnography during high-frequency ventilation in neonatal intensive care units
Amir Kugelman1,2, Ruben Bromiker3, Arieh Riskin1
1Department of Neonatology, Bnai Zion Medical Center, The B&R Rappaport Faculty of Medicine, Technion, Haifa, Israel.
Background And Objective:
High-frequency ventilation (HFV) is a powerful tool for CO2 elimination, and thus requires careful monitoring of CO2 . Our aim was to assess the diagnostic accuracy (correlation, agreement, and trending) of continuous distal capnography (dCap) with PaCO2 in infants ventilated with HFV.
Design:
This was a prospective, observational, multicenter study. dCap was compared with simultaneous PaCO2 ("gold standard") drawn from indwelling arterial line for patient care in term and preterm infants ventilated with HFV. dCap was obtained via the side-port of a double-lumen endotracheal-tube by a Microstream capnograph with specially designed software for HFV.
Results:
Twenty-four infants participated in the study (median [range] gestational age [GA]: 26.8 [23.6-38.6] weeks). Analysis included 332 measurements. dCap was in correlation (r = 0.70, P < 0.001) but with less than adequate agreement (mean difference ± SD of the differences: -11.7 ± 10.3 mmHg) with PaCO2 . Comparable findings were found in the subgroup of infants <1,000 g (n = 240 measurements). Correlations were maintained in severe lung disease. Changes in dCap and in PaCO2 for consecutive measurements within each patient were correlated (r = 0.63, P < 0.001). Area under the receiver operating curves (ROC) for dCap to detect high (>60 mmHg) or low (<30 mmHg) PaCO2 was 0.83 (CI: 0.76-0.90) and 0.88 (CI: 0.79-0.97), respectively; P < 0.001.
Conclusions:
Our prospective study suggests that continuous dCap in infants ventilated with HFV may be helpful for trends and alarm for unsafe levels of PaCO2 . dCap is only a complimentary tool and cannot replace PaCO2 sampling because the agreement between these measurements was less than adequate.
More Related Videos
Related Concept Videos
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Acute Respiratory Failure-III
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement:
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...