RETROSPECTIVE METHOD VALIDATION AND UNCERTAINTY ESTIMATION FOR ACTINIDES DETERMINATION IN EXCRETA BY ALPHA
1Bioelimination Laboratory, Radiation Dosimetry Service, CIEMAT, Avda. Complutense 40, 28040 Madrid, Spain carolina.hernandez@ciemat.es.
The Bioelimination Laboratory validated its alpha spectrometry method for quantifying internal radiation exposure in workers. This involved method validation and uncertainty estimation to meet ISO 17025 accreditation requirements.
Area of Science:
- Analytical Chemistry
- Radiochemistry
- Occupational Health
Background:
- ISO 17025 accreditation requires method validation and uncertainty estimation for testing laboratories.
- The Bioelimination Laboratory at CIEMAT quantifies alpha emitters in biological samples from radiation-exposed workers.
- Alpha spectrometry is used to measure isotopes such as Plutonium (Pu), Americium (Am), Thorium (Th), Uranium (U), and Curium (Cm).
Purpose of the Study:
- To validate the alpha spectrometry analytical method used for internal radiation monitoring.
- To estimate all sources of uncertainty affecting the quantification of alpha emitters in urine and faecal samples.
- To fulfill technical requirements for ISO 17025 accreditation.
Main Methods:
- Retrospective analysis of data obtained over several years.
- Identification and quantification of all potential sources of uncertainty.
- Calculation of the overall combined standard uncertainty for the analytical method.
Main Results:
- The analytical method for quantifying alpha emitters in biological samples was successfully validated.
- Comprehensive uncertainty estimation was performed, identifying all contributing factors.
- The overall combined standard uncertainty was calculated, providing a measure of method reliability.
Conclusions:
- The laboratory has successfully validated its alpha spectrometry method for internal radiation monitoring.
- The uncertainty estimation process provides a robust assessment of analytical results.
- These steps are crucial for achieving ISO 17025 accreditation and ensuring reliable monitoring of exposed workers.
More Related Videos
Related Concept Videos
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing...
Atomic Emission Spectroscopy: Overview
Atomic Emission Spectroscopy: Lab
Atomic Absorption Spectroscopy: Interference
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...
One-Compartment Open Model: Urinary Excretion Data and Determination of k
Atomic Absorption Spectroscopy: Overview
When irradiated by EMR of a particular wavelength, these...