Impact of incorrect filling of citrate blood sampling tubes on thromboelastometry
Thorsten Haas1, Nelly Spielmann1, Melissa Cushing2
1a Department of Anesthesia , Zurich University Children's Hospital , Zurich , Switzerland.
Background:
ROTEM(®) test results can be affected by the citrate to blood ratio in the tested sample. Overfilling or underfilling specimen tubes can change this ratio.
Objectives:
The aim of this study was to determine the impact of under- and overfilling citrate test tubes on ROTEM(®) EXTEM measurements.
Results:
Overall repeated measures ANOVA demonstrated significant differences of CT (p = 0.004), CFT (p = 0.005), A5 (p = 0.001), A10 (p <0.001), and MCF (p < 0.001). Pairwise comparison revealed that underfilling significantly altered the results of those parameters reflecting functional clot firmness (A5, A10, and MCF), while overfilling led to a prolongation of the CT results only. No differences were observed for alpha angle and maximum lysis.
Conclusion:
Both underfilling and overfilling specimen tubes have significant influence on the results of the ROTEM(®) EXTEM test, although the small observed bias is likely of limited clinical relevance. However, it seems prudent to limit the maximum allowed difference in filling to less than ± 10%. All ROTEM(®) operators should be aware of this pre-analytical variable.
Related Concept Videos
Errors occurring during blood pressure monitoring
Several factors...
Effects of EDTA on End-Point Detection Methods
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a...