Accuracy of blood pressure monitors available in high street pharmacies
Adam Ware1, Richard Stevens, Mary Selwood
1aNuffield Department of Primary Care Health Sciences bKeble College, University of Oxford, Oxford, UK.
Abstract:
The aim of this study was to assess the accuracy of automated blood pressure monitors on sale to the UK general public. We conducted static pressure accuracy testing on all compatible (19 out of 22 available) blood pressure monitors available for sale in pharmacies within the city of Oxford, UK, and tested two devices for accuracy in measurement of systolic and diastolic blood pressures in 21 adults. The devices showed good accuracy when measuring static pressure in laboratory bench testing, with the median error per device ranging from -2.2 to +1.2 mmHg; however, the two devices tested performed worse in vivo than in laboratory tests, with median errors as high as 6 mmHg. The monitors showed good accuracy in static pressure testing, with a lack of correlation between monitor price and accuracy. However, higher error rates seen during in-vivo testing of a subset of monitors may indicate that static testing may not be appropriate for routine accuracy assessment of these monitors.
More Related Videos
Related Concept Videos
Errors occurring during blood pressure monitoring
Several factors...
Measurement of Blood Pressure
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessment of blood pressure in brachial artery(two-step method)