Particle Image Velocimetry Investigation of Hemodynamics via Aortic Phantom
Published on: February 25, 2022
Aortic Flow Patterns After Simulated Implantation of Transcatheter Aortic Valves
Anastasios Kopanidis1, Ioannis Pantos, Nikolaos Alexopoulos
1Department of Mechanical Engineering, University of Western Macedonia, Kozani, Greece.
Introduction:
The functional behavior and hemodynamic characteristics of percutaneously implanted bioprosthetic valves are not known.
Methods:
We created aortic models after the simulated implantation of two of the most widely used bioprosthetic valves: the Edwards SAPIEN, and the Medtronic CoreValve. By using computational fluid dynamics analysis we sought to investigate variations in the aortic flow patterns induced by the two valve designs and their association with detrimental phenomena such as vascular remodeling, vascular wall damage and thrombosis.
Results:
The simulated implantation of models that resemble the two valves resulted in different aortic flow conditions. Vortex formation in the upper ascending aorta was more persistent in the case of the simulated Medtronic valve. The ranges of average wall shear stress (WSS) values were 2.4-3.5 Pa for Edwards and 3.0-5.3 Pa for Medtronic; the calculated WSS values induced endothelial quiescence and an atheroprotective setting in both valves. The average shear stress on the simulated valve leaflets was low; however, hotspots were present in both valves (155.0 Pa for Edwards and 250.0 Pa for Medtronic) which would in theory be able to cause platelet activation and thus promote thrombosis. The pressure drops along the aorta were slightly lower for the Edwards compared to the Medtronic valve (198.0 Pa versus 218.0 Pa).
Conclusions:
The presented method allows the assessment of aortic flow conditions following the implantation of bioprosthetic valves. It may be useful in predicting detrimental flow phenomena, thus facilitating the selection of appropriate valve designs.
More Related Videos
06:59Improved Registration of 3D CT Angiography with X-ray Fluoroscopy for Image Fusion During Transcatheter Aortic Valve Implantation
Published on: June 3, 2018
06:04Upper-extremity Approach for Secondary Access in Transfemoral Transcatheter Aortic Valve Implantation
Published on: August 8, 2025
Related Concept Videos
Aortic Regurgitation II: Clinical Features and Diagnostic Tests
Aortic Regurgitation I: Introduction