Gravimetric Analysis of Particulate Matter using Air Samplers Housing Internal Filtration Capsules
Sean O'Connor1, Paula Fey O'Connor1, H Amy Feng1
1US Department of Health and Human Services, Centers for Disease Control and Prevention, National Institute for Occupational Safety and Health, 4676 Columbia Parkway, Mail Stop R-7, Cincinnati, Ohio 45226-1998, USA, KAshley@cdc.gov.
None:
An evaluation was carried out to investigate the suitability of polyvinyl chloride (PVC) internal capsules, housed within air sampling devices, for gravimetric analysis of airborne particles collected in workplaces. Experiments were carried out using blank PVC capsules and PVC capsules spiked with 0,1 - 4 mg of National Institute of Standards and Technology Standard Reference Material® (NIST SRM) 1648 (Urban Particulate Matter) and Arizona Road Dust (Air Cleaner Test Dust). The capsules were housed within plastic closed-face cassette samplers (CFCs). A method detection limit (MDL) of 0,075 mg per sample was estimated. Precision Sr at 0,5 - 4 mg per sample was 0,031 and the estimated bias was 0,058. Weight stability over 28 days was verified for both blanks and spiked capsules. Independent laboratory testing on blanks and field samples verified long-term weight stability as well as sampling and analysis precision and bias estimates. An overall precision estimate Ŝrt of 0,059 was obtained. An accuracy measure of ±15,5% was found for the gravimetric method using PVC internal capsules.
Related Concept Videos
Gravimetry: Overview
In precipitation gravimetry, the analyte is converted into a precipitate and weighed. For example, the silver content in a sample can be estimated by precipitating and...
Sampling Methods: Sample Types
Solid samples include a variety of substances, such as sediments from water bodies, soil, metals, and biological tissues. Two standard methods for extracting sediments from water bodies are grab sampling and piston coring. Grab sampling involves using a device to collect a discrete sediment sample from the bottom of a water body with minimal disturbance. Grab samples do not always represent the entire area due to...
Precipitation Gravimetry
In determining nickel by gravimetric analysis, a precipitant of ethanolic dimethylglyoxime is added to a hot nickel salt solution. This is quickly followed by the dropwise addition of dilute ammonia solution until precipitation occurs. A...
Sample Preparation for Analysis: Overview
Bulk or large solid samples are typically reduced in size using grinding, crushing, or milling techniques to increase the...
Volatilization