Related Experiment Video
Updated: Apr 1, 2026

In Vitro Application of a Wireless Sensor in Flexion-Extension Gap Balance of Unicompartmental Knee Arthroplasty
Published on: May 5, 2023
Tibial baseplate positioning in robotic-assisted and conventional unicompartmental knee arthroplasty
Katherine P MacCallum1, Jonathan R Danoff1, Jeffrey A Geller2
1Center for Hip and Knee Replacement, Columbia University Medical Center, New York-Presbyterian Hospital, 622 West 168th Street, PH 1147, New York, NY, 10032, USA.
Abstract:
This study compared tibial baseplate alignment (TBA) between robotic-arm-assisted (RAA) and conventional (CONV) unicompartmental knee arthroplasties (UKAs). We hypothesized that RAA would increase the percentage of implants within a predetermined safe zone (SZ). We identified 177 CONV and 87 RAA UKAs through our center's patient registry. Two individuals reviewed postoperative knee radiographs and determined TBA. Coronal baseplate positioning was more accurate (i.e., within the SZ) for RAA (2.6° ± 1.5° vs. 3.9° ± 2.4°, p < 0. 0001). Conversely, sagittal alignment was more accurate for CONV (4.9° ± 2.8° vs. 2.4° ± 1.6°, p < 0.0001). RAA was more precise in both planes (p < 0.0001). There was no difference in the percentage of implants within the SZ between the two groups (p = 1.0).