Feasibility of a Modified Biopsy Needle for Irreversible Electroporation Ablation and Periprocedural Tissue Sampling
Thomas Wimmer1,2, Govindarajan Srimathveeravalli3, Mikhail Silk3
1Interventional Radiology Service, Department of Radiology, Memorial Sloan-Kettering Cancer Center, New York, NY, USA thomas.wimmer@medunigraz.at.
Objectives:
To test the feasibility of modified biopsy needles as probes for irreversible electroporation ablation and periprocedural biopsy.
Methods:
Core biopsy needles of 16-G/9-cm were customized to serve as experimental ablation probes. Computed tomography-guided percutaneous irreversible electroporation was performed in in vivo porcine kidneys with pairs of experimental (n = 10) or standard probes (n = 10) using a single parameter set (1667 V/cm, ninety 100 µs pulses). Two biopsy samples were taken immediately following ablation using the experimental probes (n = 20). Ablation outcomes were compared using computed tomography, simulation, and histology. Biopsy and necropsy histology were compared.
Results:
Simulation-suggested ablations with experimental probes were smaller than that with standard electrodes (455.23 vs 543.16 mm2), although both exhibited similar shape. Computed tomography (standard: 556 ± 61 mm2, experimental: 515 ± 67 mm2; P = .25) and histology (standard: 313 ± 77 mm2, experimental: 275 ± 75 mm2; P = .29) indicated ablations with experimental probes were not significantly different from the standard. Histopathology indicated similar morphological changes in both groups. Biopsies from the ablation zone yielded at least 1 core with sufficient tissue for analysis (11 of the 20).
Conclusions:
A combined probe for irreversible electroporation ablation and periprocedural tissue sampling from the ablation zone is feasible. Ablation outcomes are comparable to those of standard electrodes.