Background:
Six sigma provides an objective and quantitative methodology to describe the laboratory testing performance. In this study, we conducted a national trueness verification scheme with fresh frozen serum (FFS) for serum creatinine to evaluate its performance in China.
Methods:
Two different concentration levels of FFS, targeted with reference method, were sent to 98 laboratories in China. Imprecision and bias of the measurement procedure were calculated for each participant to further evaluate the sigma value. Quality goal index (QGI) analysis was used to investigate the reason of unacceptable performance for laboratories with σ < 3.
Results:
Our study indicated that the sample with high concentration of creatinine had preferable sigma values. For the enzymatic method, 7.0% (5/71) to 45.1% (32/71) of the laboratories need to improve their measurement procedures (σ < 3). And for the Jaffe method, the percentages were from 11.5% (3/26) to 73.1% (19/26). QGI analysis suggested that most of the laboratories (62.5% for the enzymatic method and 68.4% for the Jaffe method) should make an effort to improve the trueness (QGI > 1.2). Only 3.1-5.3% of the laboratories should improve both of the precision and trueness.
Conclusions:
Sigma metric analysis of the serum creatinine assays is disappointing, which was mainly due to the unacceptable analytical bias according to the QGI analysis. Further effort is needed to enhance the trueness of the creatinine measurement.
Related Concept Videos
Serum Studies: Renal Function Tests
Drug Dosing in Renal Diseases: Measurement of Serum Creatinine Concentration and Clearance
Drug Dosing in Renal Diseases: Measurement of Glomerular Filtration Rate
Drug Dosing in Renal Diseases: Estimation of Glomerular Filtration Rate Based on Serum Creatinine Concentration
Factors Affecting Renal Clearance: Renal Impairment
One condition associated with renal failure is uremia. Uremia is characterized by impaired glomerular filtration and fluid accumulation in the body. This condition hinders the renal clearance of drugs, resulting in drug accumulation and potential...
Renal Clearance
Renal clearance refers to the volume of plasma cleared of a specific substance, such as creatinine, per unit of time. To measure clearance, urine samples are collected over a 24-hour period during each bladder voiding, followed by a single blood sample at the...