MRM Microcoil Performance Calibration and Usage Demonstrated on Medicago truncatula Roots at 22 T
Published on: January 16, 2021
A novel acoustically quiet coil for neonatal MRI system
Christopher M Ireland1, Randy O Giaquinto2, Wolfgang Loew2
1Imaging Research Center, Department of Radiology, Cincinnati Children's Hospital Medical Center, 3333 Burnet Ave., Cincinnati, OH 45229, USA ; Department of Biomedical, Chemical, and Environmental Engineering, 601 Engineering Research Center, University of Cincinnati, 2901 Woodside Drive, Cincinnati, OH 45220, USA.
A new acoustically quiet MRI coil significantly reduces noise exposure for neonates, lowering sound pressure levels by an average of 9 dBA compared to conventional coils. This innovation aims to mitigate distress and developmental impacts during neonatal MRI procedures.
Area of Science:
- Medical Imaging
- Acoustics
- Neonatal Care
Background:
- Magnetic Resonance Imaging (MRI) procedures generate significant acoustic noise, posing risks of physiological distress and developmental issues for infants, particularly preterm neonates.
- High sound pressure levels during neonatal MRI can negatively impact infant well-being and require mitigation strategies.
Purpose of the Study:
- To develop and evaluate a novel acoustically quiet MRI coil designed to reduce sound pressure levels for neonates undergoing MR procedures.
- To compare the acoustic performance of the new quiet coil against a conventional body coil in a neonatal intensive care unit (NICU) MRI setting.
Main Methods:
- A new MRI coil incorporating sound-abating materials was designed with a conventional high-pass birdcage RF design.
- Acoustic performance was assessed by measuring sound pressure level and frequency response using six standard clinical MRI protocols on two small footprint NICU MRI systems.
- Comparisons were made between the acoustically quiet coil and a conventional body coil.
Main Results:
- The acoustically quiet coil demonstrated an average sound pressure level of 82.2 dBA across six protocols, compared to 91.1 dBA for the conventional body coil.
- The quiet coil was consistently 9 dBA quieter on average (range 6-10 dBA).
- While harmonic profiles were similar, the quiet coil exhibited lower acoustic amplitude, up to 20 dBA reduction.
Conclusions:
- The novel acoustically quiet MRI coil effectively reduces sound pressure levels during neonatal MR procedures.
- This coil represents a significant advancement in minimizing acoustic risks for vulnerable infant populations in the NICU.
- The findings support the use of this quieter coil technology to improve the safety and comfort of neonates undergoing MRI.