Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
Assessment of target volume doses in radiotherapy based on the standard and measured calibration curves
Gholamreza Fallah Mohammadi1, Nader Riyahi Alam, Hamed Rezaeejam
1Department of Medical Physics and Biomedical Engineering, Faculty of Medicine, Tehran University of Medical Sciences, Tehran; Department of Radiology, Faculty of Paramedical, Mazandaran University of Medical Sciences, Sari, Iran.
Context:
In radiation treatments, estimation of the dose distribution in the target volume is one of the main components of the treatment planning procedure. To estimate the dose distribution, the information of electron densities is necessary. The standard curves determined by computed tomography (CT) scanner that may be different from that of other oncology centers. In this study, the changes of dose calculation due to the different calibration curves (HU-ρel) were investigated.
Materials And Methods:
Dose values were calculated based on the standard calibration curve that was predefined for the treatment planning system (TPS). The calibration curve was also extracted from the CT images of the phantom, and dose values were calculated based on this curve. The percentage errors of the calculated values were determined.
Statistical Analysis Used:
The statistical analyses of the mean differences were performed using the Wilcoxon rank-sum test for both of the calibration curves.
Results And Discussion:
The results show no significant difference for both of the measured and standard calibration curves (HU-ρel) in 6, 15, and 18 MeV energies. In Wilcoxon ranked sum nonparametric test for independent samples with P<0.05, the equality of monitor units for both of the curves to transfer 200 cGy doses to reference points was resulted. The percentage errors of the calculated values were lower than 2% and 1.5% in 6 and 15 MeV, respectively.
Conclusion:
From the results, it could be concluded that the standard calibration curve could be used in TPS dose calculation accurately.
Related Concept Videos
Calibration Curves: Linear Least Squares
For data that follow a straight line, the standard method for fitting is the linear...
Glassware Calibration
Volumetric flasks: Volumetric flasks are designed to prepare aqueous solutions of precise volumes accurately with a calibration line on the neck. To calibrate a volumetric flask, it is important to fill it with distilled...
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Dose Response Curve: Conventional Versus Nonmonotonic
Dose Size and Dosing Frequency: Determination Methods
Calibration Curves: Correlation Coefficient