Relative contribution of biological variation and technical variables to zone diameter variations of disc diffusion
Michael Hombach1, Carlos Ochoa2, Florian P Maurer3
1Institut für Medizinische Mikrobiologie, Universität Zürich, 8006 Zurich, Switzerland mhombach@imm.uzh.ch.
Objectives:
Disc diffusion is still largely based on manual procedures. Technical variations originate from inoculum preparation, variations in materials, individual operator plate streaking and reading accuracy. Resulting measurement imprecision contributes to categorization errors. Biological variation resembles the natural fluctuation of a measured parameter such as antibiotic susceptibility around a mean value. It is deemed to originate from factors such as genetic background or metabolic state. This study analysed the relative contribution of different technical and biological factors to total disc diffusion variation.
Methods:
For calculation of relative error factor contribution to disc diffusion variability, five experiments were designed keeping different combinations of error factors constant. A mathematical model was developed to analyse the individual error factor contribution to disc diffusion variation for each of the tested drug-species combinations.
Results:
The contribution of biological variation to total diameter variance ranged from 10.4% to 98.8% for different drug-species combinations. Highest biological variation was found for Enterococcus faecalis WT and vancomycin (98.8%) and for penicillinase-producing Staphylococcus aureus and penicillin G (96.0%). Average imprecision of automated zone reading revealed that 1.4%-5.3% of total imprecision was due to technical variation, while materials, i.e. antibiotic discs and agar plates, contributed between 2.6% and 3.9%. Inoculum preparation and manual plate streaking contributed 6.8%-24.8% and 6.6%-24.3%, respectively, to total imprecision.
Conclusions:
This study illustrates the relative contributions of technical factors that account for a significant part of total variance in disc diffusion. The highest relative contribution originated from the operator, i.e. manual inoculum preparation and plate streaking. Further standardization of inoculum preparation and plate streaking by automation could potentially increase the precision of disc diffusion and improve the correlation of susceptibility reports with clinical outcome.
More Related Videos
09:15Enhanced Reproducibility and Precision of High-Throughput Quantification of Bacterial Growth Data Using a Microplate Reader
Published on: July 27, 2022
07:42Deferred Growth Inhibition Assay to Quantify the Effect of Bacteria-derived Antimicrobials on Competition
Published on: September 3, 2016
Related Concept Videos
Variability: Analysis
The range is a simple measure of variability, indicating the difference between the highest and...
Variation
When independent and dependent variables are plotted on a scatter plot, the slope of a line is a value that describes the rate of change between the two...