Related Experiment Video
Updated: Mar 31, 2026

Pore-scale Imaging and Characterization of Hydrocarbon Reservoir Rock Wettability at Subsurface Conditions Using X-ray Microtomography
Published on: October 21, 2018
Computed Tomography Artifact Created by Air in the X-ray Tube Oil
Wayne R Hedrick1, Michael A Markovic, James H Short
1From the Radiology Department, Aultman Hospital, and Northeast Ohio Medical University, Rootstown, OH.
Abstract:
A subtle artifact of patchy hypodensities in computed tomography images of the head mimicked acute or subacute cerebral infarct. The cause of the artifact was air in the oil of the x-ray tube. The artifact manifested only when the acquisition parameters included a rotation time of 0.5 second and a gantry tilt angle of 11 to 20 degrees. Routine quality control testing did not detect nonuniformities in the water phantom.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
X-ray Imaging
Imaging Studies III: Computed Tomography