Effect of using a Planecta™ port with a three-way stopcock on the natural frequency of blood pressure transducer kits
Shigeki Fujiwara1, Keiichi Tachihara2, Satoshi Mori2
1Department of Anesthesiology, Graduate School of Dentistry, Kanagawa Dental University, 82 Inaoka-cho, Yokosuka-shi, Kanagawa, 238-0003, Japan. josephluke@dance.ocn.ne.jp.
Abstract:
Blood pressure transducer kits are equipped with two types of Planecta™ ports-the flat-type Planecta™ port (FTP) and the Planecta™ port with a three-way stopcock (PTS). We reported that FTP application decreased the natural frequency of the kits. However, Planecta™ is an invaluable tool as it prevents infection, ensures technical simplicity, and excludes air. Hence, an ideal Planecta™ port that does not decrease the frequency characteristics is required. As a first step in this direction, we aimed to assess the influence of PTSs on the natural frequency of blood transducer kits. A DTXplus transducer kit (DT4812J; Argon Medical Devices, TX, USA) was used along with ≥1 PTSs (JMS, Hiroshima, Japan), and the frequency characteristics were assessed. The natural frequency and damping coefficient of each kit were obtained by using frequency characteristics analysis software, and these parameters were evaluated by plotting them on Gardner's chart. Regardless of whether one or two PTSs were inserted, the natural frequency of the kits only slightly decreased (from 42.5 to 41.1 Hz, when 2 PTSs were used). Thus, the frequency characteristics of the kits with PTSs were adequate for pressure monitoring. The insertion of ≥2 FTPs in pressure transducer kits should be avoided, as they markedly decrease the natural frequency and lead to underdamping. However, the effect of PTS insertion in pressure transducer kits on the frequency characteristics is minimal. Thus, we found that the use of PTS markedly improved the frequency characteristics as compared to the use of FTP.
More Related Videos
Related Concept Videos
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Measurement of Blood Pressure
Assessment of blood pressure in brachial artery(two-step method)
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Measurement of Fluid Pressure
A basic form of manometer is the piezometer, a vertical tube open at the top and filled with the same...