The novel active fixation coronary sinus lead: efficacy and safety of transvenous extraction procedure

Luca Bontempi1, Francesca Vassanelli2, Najat Ashofair1

  • 1Chair and Unit of Cardiology, University of Brescia, Spedali Civili Hospital, Brescia, Italy.

Insights

Lead stability is a challenge in cardiac resynchronization therapy (CRT). A new active fixation left ventricular lead was successfully and safely extracted, demonstrating a viable solution for lead management in CRT patients.

Area of Science:

  • Cardiology
  • Medical Device Technology

Background:

  • Cardiac resynchronization therapy (CRT) improves heart failure outcomes.
  • Complications with coronary sinus (CS) leads, such as dislodgement and high pacing thresholds, negatively impact patient quality of life and healthcare costs.
  • Lead stability remains a significant unmet need in CRT procedures.

Observation:

  • A novel active fixation left ventricular (LV) lead (Medtronic 20066 Attain Stability®) features a small side helix for secure vein placement.
  • This study reports the first human experience with the extraction of this specific active fixation LV lead.

Findings:

  • The extraction of the novel active fixation LV lead was performed safely and effectively.
  • No dissection or damage to the CS was observed during or after lead removal.
  • A rotation maneuver combined with moderate traction proved effective for lead extraction.

Implications:

  • This extraction technique offers a safe and effective solution for managing novel active fixation LV leads.
  • Successful lead extraction contributes to improved patient outcomes and potentially reduced healthcare burdens associated with CRT complications.
  • Further research should validate these findings in larger patient cohorts.