Tracheal tube biofilm removal through a novel closed-suctioning system: an experimental study
E Aguilera Xiol1, G Li Bassi2, D Wyncoll3
1Department of Pulmonary and Critical Care Medicine, Division of Animal Experimentation, Thorax Institute, Hospital Clínic, Barcelona, Spain Institut d'Investigacions Biomèdiques August Pi i Sunyer (IDIBAPS), Barcelona, Spain Centro de Investigación Biomédica en Red Enfermedades Respiratorias, Mallorca, Spain.
Background:
Tracheal tube biofilm develops during mechanical ventilation. We compared a novel closed-suctioning system vs standard closed-suctioning system in the prevention of tracheal tube biofilm.
Methods:
Eighteen pigs, on mechanical ventilation for 76 h, with P. aeruginosa pneumonia were randomized to be tracheally suctioned via the KIMVENT* closed-suctioning system (control group) or a novel closed-suctioning system (treatment group), designed to remove tracheal tube biofilm through saline jets and an inflatable balloon. Upon autopsy, two tracheal tube hemi-sections were dissected for confocal and scanning electron microscopy. Biofilm area, maximal and minimal thickness were computed. Biofilm stage was assessed.
Results:
Sixteen animals were included in the final analysis. In the treatment and control group, the mean (sd) pulmonary burden was 3.34 (1.28) and 4.17 (1.09) log cfu gr(-1), respectively (P=0.18). Tracheal tube P. aeruginosa colonization was 5.6 (4.9-6.3) and 6.2 (5.6-6.9) cfu ml(-1) (median and interquartile range) in the treatment and control group, respectively (P=0.23). In the treatment group, median biofilm area was 3.65 (3.22-4.21) log10 μm2 compared with 4.49 (4.27-4.52) log10 μm2 in the control group (P=0.031). In the treatment and control groups, the maximal biofilm thickness was 48.3 (26.7-71.2) µm (median and interquartile range) and 88.8 (43.8-125.7) µm, respectively. The minimal thickness in the treatment and control group was 0.6 (0-4.0) µm and 23.7 (5.3-27.8) µm (P=0.040) (P=0.017). Earlier stages of biofilm development were found in the treatment group (P<0.001).
Conclusions:
The novel CSS reduces biofilm accumulation within the tracheal tube. A clinical trial is required to confirm these findings and the impact on major outcomes.
Related Concept Videos
Tracheostomy Suctioning II: Procedure
Tracheostomy Suctioning I: Pre-Procedural Steps
Equipment Required
First, gather all necessary equipment: a sterile suction catheter, a sterile disposable container, sterile gloves, a towel or...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Tracheostomy: Procedure and Tubes
Tracheostomy tubes can be made of semiflexible plastic (polyurethane or silicone), rigid plastic, or metal, and they come in...
Suctioning the Nasopharyngeal Airway
Equipment Required
Endotracheal Tube Extubation
Procedure
Extubation removes the endotracheal tube (ETT) from the patient on mechanical ventilation. It requires a well-coordinated, multidisciplinary approach involving physicians, nurses, respiratory therapists, and other healthcare professionals....