Related Experiment Video
Updated: Mar 31, 2026

Monitoring Lung Function with Electrical Impedance Tomography in the Intensive Care Unit
Published on: September 6, 2024
Effect of Trigger Sensitivity on Redistribution of Ventilation During Pressure Support Ventilation Detected by
Oliver C Radke1, Thomas Schneider2, Elisabeth Vogel2
1Department of Anesthesiology and Intensive Care Medicine,Klinikum Bremerhaven-Reinkenheide, Postbrookstr. 103, 27574 Bremerhaven, Germany ; Department of Anesthesia and Perioperative Care, San Francisco General Hospital, University of California San Francisco, San Francisco, California, USA ; Department of Anaesthesiology and Intensive Care Medicine, University Hospital, Technische Universitat Dresden, Germany.
Background:
In supine position, pressure support ventilation causes a redistribution of ventilation towards the ventral regions of the lung. Theoretically, a less sensitive support trigger would cause the patient to breathe more actively, potentially attenuating the effect of positive pressure ventilation.
Objectives:
To quantify the effect of trigger setting, we assessed redistribution of ventilation during pressure support ventilation (PSV) using electrical impedance tomography (EIT).
Patients And Methods:
With approval from the local ethics committee, six orthopedic patients were enrolled. All patients had general anesthesia with a laryngeal mask airway and a standardized anesthetic regimen (sufentanil, propofol and sevoflurane). Pressure support trigger settings varied between 2 and 15 L/minute and compared to unassisted spontaneous breathing. From EIT data, the center of ventilation (COV), the fraction of the total ventilation per region of interest (ROI) and intratidal gas distribution were calculated.
Results:
At all trigger settings, pressure support ventilation caused a significant ventral shift of the center of ventilation compared with during spontaneous breathing, confirmed by the analysis by regions of interest. During spontaneous breathing, COV was not different from baseline values obtained before induction of anesthesia. During PSV, the intratidal regional gas distribution (ITV-analysis) revealed subtle changes during the early inspiratory phase not detected by the COV-analysis.
Conclusions:
Pressure support ventilation, but not spontaneous breathing, induces a significant redistribution of ventilation towards the ventral region. The sensitivity of the support trigger appears to influence the distribution of ventilation only during the early phase of inspiration.
More Related Videos
08:26Quantitative Mapping of Specific Ventilation in the Human Lung using Proton Magnetic Resonance Imaging and Oxygen as a Contrast Agent
Published on: June 5, 2019
09:08Author Spotlight: Standardization and Best Practices for Advancing Lung Imaging Using 129Xe MRI
Published on: November 21, 2023
Related Concept Videos
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement:
Mechanical Ventilation I: Indication and Settings
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Factors Affecting Pulmonary Ventilation
Alveolar Surface Tension
The alveolar fluid lines the luminal surface of the alveoli and exerts a force called surface tension. This force is caused by the polar water molecules in the liquid being more strongly attracted to each...