Evaluation of the Ambu CPAP system.

P J Venn1

  • 1Nuffield Department of Anaesthetics, Radcliffe Infirmary, Oxford.

British Journal of Anaesthesia
|March 1, 1989
PubMed
Summary

This study evaluated a cost-effective Venturi-driven device for continuous positive airway pressure (CPAP) delivery. The system demonstrated good performance and low gas consumption, suitable for spontaneous breathing applications.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Quality care in anaesthesia: roles of regulation and accreditation.

British journal of anaesthesia·2015
Same author

Peri-operative management of the obese surgical patient 2015: Association of Anaesthetists of Great Britain and Ireland Society for Obesity and Bariatric Anaesthesia.

Anaesthesia·2015
Same author

The influence of timing and route of administration of intravenous ketorolac on analgesia after hand surgery.

Anaesthesia·2002
Same author

Management of Jehovah's Witness patients undergoing major head and neck surgery.

Head & neck·1999
Same author

Use of laryngeal mask during tonsillectomy.

British journal of anaesthesia·1998
Same author

Fibreoptic intubation through a laryngeal mask airway in an infant with Robin sequence.

European journal of anaesthesiology·1998

Area of Science:

  • Respiratory Medicine
  • Biomedical Engineering
  • Medical Device Technology

Background:

  • Continuous Positive Airway Pressure (CPAP) is crucial for respiratory support.
  • Existing CPAP delivery systems vary in cost and efficiency.
  • Venturi-driven devices offer a potential alternative for CPAP delivery.

Purpose of the Study:

  • To assess the performance of a Venturi-driven CPAP delivery device.
  • To evaluate its efficiency and cost-effectiveness under simulated breathing conditions.
  • To characterize the behavior of its components, including reservoir bags and expiratory valves.

Main Methods:

  • A Venturi-driven CPAP system with a reservoir bag and expiratory valve was tested.
  • Simulated spontaneous breathing conditions were employed.
  • System performance was evaluated based on gas consumption, oxygen delivery, and pressure characteristics.

Main Results:

  • The device was economical, using only 3.5 L/min of fresh gas.
  • One Venturi device showed inadequate performance due to partial blockage.
  • A second device closely met manufacturer specifications for oxygen and flow.
  • Reservoir bag compliance increased with circuit pressure, and the expiratory valve functioned as a threshold resistor.
  • Airway pressure monitoring overestimated true pressure by 20% above 1 kPa.

Conclusions:

  • The Venturi-driven CPAP system is a potentially cost-effective and efficient option for respiratory support.
  • Careful calibration and monitoring are necessary due to potential Venturi blockages and pressure measurement inaccuracies.
  • Further research may optimize this system for clinical application.

Related Experiment Videos