Multicenter Clinical Trials Using 18F-FDG PET to Measure Early Response to Oncologic Therapy: Effects of Injection-to-Acquisition Time Variability on Required Sample Size
Related Experiment Video
Updated: Mar 31, 2026

Gene Regulation and Targeted Therapy in Gastric Cancer Peritoneal Metastasis: Radiological Findings from Dual Energy CT and PET/CT
Published on: January 22, 2018
Brenda F Kurland1, Mark Muzi2, Lanell M Peterson2
1Department of Biostatistics, University of Pittsburgh, Pittsburgh, Pennsylvania bfk10@pitt.edu.
Unlabelled:
Uptake time (interval between tracer injection and image acquisition) affects the SUV measured for tumors in (18)F-FDG PET images. With dissimilar uptake times, changes in tumor SUVs will be under- or overestimated. This study examined the influence of uptake time on tumor response assessment using a virtual clinical trials approach.
Methods:
Tumor kinetic parameters were estimated from dynamic (18)F-FDG PET scans of breast cancer patients and used to simulate time-activity curves for 45-120 min after injection. Five-minute uptake time frames followed 4 scenarios: the first was a standardized static uptake time (the SUV from 60 to 65 min was selected for all scans), the second was uptake times sampled from an academic PET facility with strict adherence to standardization protocols, the third was a distribution similar to scenario 2 but with greater deviation from standards, and the fourth was a mixture of hurried scans (45- to 65-min start of image acquisition) and frequent delays (58- to 115-min uptake time). The proportion of out-of-range scans (<50 or >70 min, or >15-min difference between paired scans) was 0%, 20%, 44%, and 64% for scenarios 1, 2, 3, and 4, respectively. A published SUV correction based on local linearity of uptake-time dependence was applied in a separate analysis. Influence of uptake-time variation was assessed as sensitivity for detecting response (probability of observing a change of ≥30% decrease in (18)F-FDG PET SUV given a true decrease of 40%) and specificity (probability of observing an absolute change of <30% given no true change).
Results:
Sensitivity was 96% for scenario 1, and ranged from 73% for scenario 4 (95% confidence interval, 70%-76%) to 92% (90%-93%) for scenario 2. Specificity for all scenarios was at least 91%. Single-arm phase II trials required an 8%-115% greater sample size for scenarios 2-4 than for scenario 1. If uptake time is known, SUV correction methods may raise sensitivity to 87%-95% and reduce the sample size increase to less than 27%.
Conclusion:
Uptake-time deviations from standardized protocols occur frequently, potentially decreasing the performance of (18)F-FDG PET response biomarkers. Correcting SUV for uptake time improves sensitivity, but algorithm refinement is needed. Stricter uptake-time control and effective correction algorithms could improve power and decrease costs for clinical trials using (18)F-FDG PET endpoints.
More Related Videos
06:51Utilizing 18F-FDG PET/CT Imaging and Quantitative Histology to Measure Dynamic Changes in the Glucose Metabolism in Mouse Models of Lung Cancer
Published on: July 21, 2018
10:33Automated Preparation of [68Ga]Ga-3BP-3940 on a Synthesis Module for PET Imaging of the Tumor Microenvironment
Published on: April 25, 2025
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...