Multicenter analytical performance evaluation of a fully automated anti-Müllerian hormone assay and reference
E Anckaert1, M Öktem2, A Thies3
1Laboratory of Hormonology and Tumour Markers, Universitair Ziekenhuis Brussel, Free University of Brussels (VUB), Brussels, Belgium.
Objectives:
Anti-Müllenria hormone (AMH) is an established biomarker for assessing ovarian reserve and predicting response to controlled ovarian stimulation. Its routine clinical use is hampered by the variability and low-throughput of available enzyme-linked immunosorbent assays (ELISA). The presented study examined if a fully automated AMH electrochemiluminescence assay (ECLIA; Elecsys® AMH assay, Roche Diagnostics) was suitable for measuring AMH levels in healthy women and in those diagnosed with polycystic ovary syndrome (PCOS).
Design And Methods:
Five European laboratories evaluated the Elecsys® AMH assay independently under routine conditions over eight months. Within-run imprecision, repeatability, intermediate precision, linearity and functional sensitivity were assessed. The Elecsys® AMH assay was compared to a manual ELISA microtiter plate format test (AMH Gen II ELISA, modified version; Beckman Coulter Inc.) using 1729 routine serum samples. AMH reference intervals were determined in 887 healthy women with regular menstrual cycle aged 20–50 years, and 149 women diagnosed with PCOS.
Results:
The fully automated Elecsys® AMH assay showed excellent precision, linearity, and functional sensitivity. The coefficient of variation was 1.8% for repeatability and 4.4% for intermediate precision. Values measured with the Elecsys® AMH assay were highly correlated with the manual ELISA method (modified version) but 24–28% lower. Reference intervals showed the expected AMH decline with age in healthy women and increased AMH levels in women with PCOS.
Conclusions:
The Elecsys® AMH assay demonstrated good precision under routine conditions, and is suitable for determining AMH levels in serum and lithium-heparin plasma.