Performance characteristics of mobile MOSFET dosimeter for kilovoltage X-rays used in image guided radiotherapy
A Sathish Kumar1, I Rabi Raja Singh1, S D Sharma2
1Department of Radiotherapy, Christian Medical College, Vellore, Tamil Nadu, India.
Abstract:
The main objective of this study was to investigate the characteristics of metal oxide semiconductor field effect transistor (MOSFET) dosimeter for kilovoltage (kV) X-ray beams in order to perform the in vivo dosimetry during image guidance in radiotherapy. The performance characteristics of high sensitivity MOSFET dosimeters were investigated for 80, 90, 100, 110, 120, and 125 kV X-ray beams used for imaging in radiotherapy. This study was performed using Clinac 2100 C/D medical electron linear accelerator with on-board imaging and kV cone beam computed tomography system. The characteristics studied in this work include energy dependence, angular dependence, and linearity. The X-ray beam outputs were measured as per American Association of Physicists in Medicine (AAPM) TG 61 recommendations using PTW parallel plate (PP) ionization chamber, which was calibrated in terms of air kerma (Nk) by the National Standard Laboratory. The MOSFET dosimeters were calibrated against the PP ionization chamber for all the kV X-ray beams and the calibration coefficient was found to be 0.11 cGy/mV with a standard deviation of about ±1%. The response of MOSFET was found to be energy independent for the kV X-ray energies used in this study. The response of the MOSFET dosimeter was also found independent of angle of incidence for the gantry angles in the range of 0° to 360° in-air as well as at 3 cm depth in tissue equivalent phantom.
More Related Videos
06:21Diffuse Optical Spectroscopy for the Quantitative Assessment of Acute Ionizing Radiation Induced Skin Toxicity Using a Mouse Model
Published on: May 27, 2016
07:31Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
Related Concept Videos
Characteristics of MOSFET
Various vital parameters influence their functionality, which is crucial for theory and electronics applications. First, channel dimensions, precisely length, and width, are pivotal. The size of these channels affects the transistor's ability to carry current and switching speeds; shorter channels typically enable...
MOSFET: Depletion Mode
The primary characteristic of depletion-mode MOSFETs is their ability to conduct current between the drain and source terminals without gate bias. This inherent conductivity...
MOSFET
In an n-MOSFET, the structure includes n-type source and drain...
Biological Effects of Radiation