Image rejects in general direct digital radiography
Bjørn Hofmann1, Tine Blomberg Rosanowsky2, Camilla Jensen2
1Section for Health, Technology, and Society, Gjøvik University College, Gjøvik, Norway and Centre for Medical Ethics, University of Oslo, Oslo, Norway.
Digital radiography still has high image rejection rates, comparable to film systems. Most deleted images result from positioning or centering errors, indicating a need for improved training and quality control in radiology departments.
Area of Science:
- Radiology
- Medical Imaging
- Quality Improvement
Background:
- Image reject analysis is crucial for assessing image quality and identifying unnecessary procedures in radiology.
- While common in the film era, reject analysis has seen fewer studies since the transition to digital radiography.
- A misconception exists that digitalization has eliminated the issue of image rejects.
Purpose of the Study:
- To quantify the rate of deleted images in direct digital radiography (DR).
- To evaluate the extent of image rejects and unnecessary imaging in DR.
- To identify the causes of image deletions to enhance radiological services.
Main Methods:
- Review of all exposed images from two direct digital radiography laboratories over January 2014.
- Registration of examination types, total exposed images, and deleted images.
- Individual analysis of each deleted image to determine the specific reason for deletion.
Main Results:
- A deletion rate of 11% was observed, with 596 out of 5417 exposed images being deleted.
- Positioning errors accounted for 51.3% of deletions, followed by centering errors at 31.0%.
- Highest deletion rates were found in knee (20.6%), hip (18.5%), and ankle (13.8%) examinations.
Conclusions:
- The reject rate in digital radiography is as high as in previous film-based systems.
- Reasons for image rejection differ significantly between digital and film systems, challenging the assumption that digitalization eliminates rejects.
- High rates of deleted (unnecessary) images necessitate improvements in management, training, and quality assurance within radiology departments.
Related Concept Videos
X-ray Imaging
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...