[Determination of lead in workplace air by portable X-ray fluorescence spectrometry]
Objective:
To investigate the permance index ofof portable X-ray fluorescence spectrometer in the determination of lead on filter membrane and to provide data for the determination of lead in workplace air.
Methods:
Irradiated with X-ray, the lead would emit specific X-ray fluorescence during the process from the excited state back to the ground state. Rapid determination of lead was completed using fluorescence energy and wave length for qualitative analysis and fluorescence intensity for quantitative measurement. Under set conditions, a series of customized calibration samples were measured to create a standard curve for quantitative analysis of lead on filter membrane.
Results:
The regression equation obtained using a portable X-ray fluorescence spectrometer to determine the lead on filter membrane was y=0.004x-0.182 (r2= 0.9999). The linear range was 0.00 -10.40 mg/m3, the minimum detectable concentration was 0.53 µg/m3, and the minimum quantifiable concentration was 1.76µg/m3. The relative standard deviation (RSD) of within-run precision of samples with different concentrations was 0.48%-6.22%, the RSD of between-run precision was 2.51%-5.09%, and the degree of accuracy was in the calibration range of standard samples.
Conclusion:
Portable X-ray fluorescence spectrometry is a simple, rapid, repeatable, and accurate method for the determination of lead on filter membrane.
Related Concept Videos
Atomic Emission Spectroscopy: Lab
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing...
Atomic Emission Spectroscopy: Overview
Atomic Fluorescence Spectroscopy
Gas Chromatography: Types of Detectors-II
Atomic Absorption Spectroscopy: Radiation and Light Sources
Two common narrow-range 'line' sources used in AAS are hollow-cathode lamps (HCLs) and...