Quantitative Detection of Trace Explosive Vapors by Programmed Temperature Desorption Gas Chromatography-Electron Capture Detector
Published on: July 25, 2014
[Determination of epichlorohydrin in workplace air by gas chromatograph-electron capture detector]
Objective:
To develop a method for determining epichlorohydrin in the workplace air by gas chromatograph-electron capture detector (GC-ECD).
Methods:
Epichlorohydrin in the workplace air was collected by activated charcoal tubes, desorbed using acetone, and analyzed by GC-ECD.
Results:
A good linearity was obtained in the range of 1.0-50 µg/mL (r=0.999 7). The detection limit was 0.012 µg/ml, while the recovery rate was 88.1% and relative standard deviation ranged from 1.11% to 3.57%. The samples could be stored for seven days at room temperature.
Conclusion:
This method effectively eliminates the interferences of alkanes on determination of epichlorohydrin and improves the sensitivity by 1 to 2 orders of magnitude, which can solve the problem of detection limit above standard in GBZ/T 160.58-2004.
Related Concept Videos
Gas Chromatography: Types of Detectors-II
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a...
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall....
High-Performance Liquid Chromatography: Types of Detectors