Influence of Sampling Procedure on Codeine Concentrations in Oral Fluid
Line D Coucke1, Lien De Smet2, Alain G Verstraete3
1Department of Laboratory Medicine, Ghent University Hospital, Ghent, Belgium.
The Saliva Collection System (SCS) demonstrated a better correlation between oral fluid (OF) and plasma codeine concentrations compared to Quantisal. This suggests SCS may offer more reliable drug monitoring via oral fluid analysis.
Area of Science:
- Pharmacology
- Analytical Chemistry
- Forensic Toxicology
Background:
- Oral fluid (OF) drug analysis faces challenges due to variable OF pH, oral contamination, and sample volume.
- Accurate correlation between plasma and OF drug concentrations is crucial for therapeutic drug monitoring and forensic applications.
Purpose of the Study:
- To compare the oral fluid (OF) to plasma ratio and concentration variability of codeine using two collection systems: Saliva Collection System (SCS) and Quantisal.
- To evaluate the reliability of OF drug measurements for predicting plasma concentrations.
Main Methods:
- 12 volunteers ingested codeine phosphate (19.5 mg).
- Oral fluid (OF) and plasma samples were collected.
- OF was collected using either SCS or Quantisal devices.
- Codeine concentrations were analyzed using gas chromatography-mass spectrometry.
Main Results:
- Mean codeine concentration at 1 hour was 29.8 μg/L in plasma, 72.8 μg/L in SCS OF, and 85.3 μg/L in Quantisal OF.
- The OF/plasma ratio was 2.30 (SCS) and 2.69 (Quantisal).
- Pearson's correlation coefficient (R²) was significantly higher for SCS (0.745) than Quantisal (0.403).
Conclusions:
- The Saliva Collection System (SCS) shows a stronger correlation between oral fluid and plasma codeine concentrations compared to Quantisal.
- SCS appears to be a more reliable method for oral fluid drug concentration analysis, especially for codeine.
- The order of sample collection influenced Quantisal's variability, with reduced variability when used after SCS.
More Related Videos
13:35A Convenient Method for Extraction and Analysis with High-Pressure Liquid Chromatography of Catecholamine Neurotransmitters and Their Metabolites
Published on: March 1, 2018
10:13Author Spotlight: An Efficient Methodology to Confidently Differentiate and Characterize Fentanyl Analogs
Published on: November 8, 2024
Related Concept Videos
Drug Concentrations: Measurements
Plasma...
Drug Concentration Versus Time Correlation
Two pivotal parameters are the minimum effective concentration (MEC) and the minimum toxic concentration (MTC). The MEC is the...
Time Course of Drug Effect
Dosage Regimen: Multiple Oral Dosage
Factors Influencing Drug Absorption: Presystemic Elimination
IV Infusion to Oral Dosing: Conversion Methods