Iatrogenic embolization and transcatheter retrieval of a ventricular septal defect occluder in a dog
Jaylyn A Durham1, Brian A Scansen1, John D Bonagura1
1Department of Clinical Sciences, The Ohio State University College of Veterinary Medicine, 601 Vernon L Tharp St., Columbus, OH 43210, USA.
Abstract:
A 7-month-old Irish Setter underwent transcatheter therapy of a muscular ventricular septal defect (VSD) and pulmonary valve stenosis. Standard devices for muscular VSD closure could not span the interventricular septum due to right ventricular hypertrophy, and an Amplatzer post-infarction muscular VSD occluder with a wider waist was successfully implanted. Following VSD closure, inflation of the balloon dilation catheter during balloon pulmonary valvuloplasty resulted in iatrogenic embolization of the VSD occluder to the left ventricular outflow tract. Retrieval and reimplantation of the device was achieved using a snare catheter. This report describes a potential complication and management during intracardiac device implantation in a dog. Additionally, the case illustrates that the Amplatzer post-infarction muscular VSD occluder holds potential value in animals with a hypertrophied interventricular septum that cannot be spanned using a conventional device.
More Related Videos
05:52Myocardial Infarction by Percutaneous Embolization Coil Deployment in a Swine Model
Published on: November 4, 2021
08:00Reduction of Iatrogenic Atrial Septal Defects with an Anterior and Inferior Transseptal Puncture Site when Operating the Cryoballoon Ablation Catheter
Published on: June 15, 2015