Pulse Oximetry Screening to Detect Cyanotic Congenital Heart Disease in Sick Neonates in a Neonatal Intensive Care
N B Mathur1, A Gupta, S Kurien
1Department of Neonatology, Maulana Azad Medical College; and *Department of Cardiology, GB Pant Hospital; New Delhi, India. Correspondence to: Dr NB Mathur, Director Professor and Head, Department of Neonatology, Maulana Azad Medical College, New Delhi, India. drnbmathur@gmail.com.
Objective:
To evaluate pulse oximetry for detection of congenital cyanotic heart disease in sick neonates using echocardiography as gold standard.
Methods:
Pulse oximetry readings were taken at admission from 950 neonates from right upper limb and either foot with infant breathing room air. Pulse oximetry was considered abnormal if oxygen saturation at room air measured <90% or difference between right hand and foot was more than 3%. Persistent abnormality was considered positive result. Echocardiography was performed on all neonates with positive pulse oximetry (study group) and on one subsequent neonate with negative screen for each neonate with positive screen (controls).
Results:
Pulse oximetry was positive in 210 neonates. It detected 20 out of 21 (95.2%) true positives. The sensitivity, specificity, positive predictive value, negative predictive value and odds ratio (95% CI) of pulse oximetry was 95.2%, 52.4%, 9.5, 99.5 and 22 (5.3, 91.4), respectively.
Conclusion:
Pulse oximetry screening is useful in detecting cyanotic heart diseases in sick newborns.
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Respiratory Assessment: Purpose and Indications
Objectives and Importance:
The primary goal of respiratory assessment is to evaluate patients at early risk of clinical deterioration. Since respiratory distress often precedes other signs of declining health, breathing patterns and sounds become a...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...