Correction of quantification errors in pelvic and spinal lesions caused by ignoring higher photon attenuation of bone

Georg Schramm1, Jens Maus1, Frank Hofheinz1

  • 1Helmholtz-Zentrum Dresden-Rossendorf, Institute of Radiopharmaceutical Cancer Research, Dresden 01328, Germany.

Medical Physics
|November 2, 2015
PubMed
Abstract

Insights

Standard MRI-based attenuation correction (MRAC) in PET/MRI underestimates bone attenuation. A new method accurately corrects for bone attenuation in spinal and pelvic lesions, improving SUV quantification for [18F]NaF PET/MRI scans.

Area of Science:

  • Medical Imaging
  • Nuclear Medicine
  • Radiology

Background:

  • Standard MRI-based attenuation correction (MRAC) in PET/MRI often overlooks bone's higher attenuation due to signal limitations.
  • This leads to inaccuracies in Standardized Uptake Value (SUV) quantification, particularly for bone lesions.

Purpose of the Study:

  • To quantify the bias introduced by standard MRAC (MRACnobone) in spinal and pelvic lesions using [18F]NaF PET/MRI.
  • To develop and validate an improved MRAC method (MRACbone) that incorporates bone attenuation.

Main Methods:

  • Reconstructed 20 [18F]NaF PET/MRI datasets using both standard MRACnobone and a novel MRACbone method with automatic bone segmentation.
  • Validated bone segmentation against PET/CT data.
  • Calculated SUV differences for spinal/pelvic lesions and other organs (lung, liver, bladder).

Main Results:

  • The novel bone segmentation showed good accuracy in the spine and pelvis.
  • MRACnobone underestimated SUVmax in pelvic and spinal lesions by -8.8% and -8.1% respectively (p<0.01).
  • Maximum SUVmax underestimation reached -13.7% in a cervical spine lesion; differences in non-bone organs were <3%.

Conclusions:

  • The developed automatic [18F]NaF PET-based bone segmentation effectively incorporates higher bone attenuation into whole-body MRAC.
  • This improves SUV quantification accuracy for pelvic and spinal lesions in [18F]NaF PET/MRI.
  • Standard MRAC provides acceptable accuracy for non-bone structures.