Related Experiment Video
Updated: Mar 30, 2026

Split Point Analysis and Uncertainty Quantification of Thermal-Optical Organic/Elemental Carbon Measurements
Published on: September 7, 2019
A method for monitoring mass concentration of black carbon particulate matter using photothermal interferometry
Baosheng Li1, Yicheng Wang2, Zhengqiang Li3
1School of Instrument Sciences and Opto-electronics Engineering, Hefei University of Technology, Hefei, 230009, China. libaosheng@hfut.edu.cn.
Abstract:
A method for measurements of mass concentration of black carbon particulate matter (PM) is proposed based on photothermal interferometry (PTI). A folded Jamin photothermal interferometer was used with a laser irradiation of particles deposited on a filter paper. The black carbon PM deposited on the filter paper was regarded as a film while the quartz filter paper was regarded as a substrate to establish a mathematical model for measuring the mass concentration of PM using a photothermal method. The photothermal interferometry system was calibrated and used to measure the atmospheric PM concentration corresponding to different dust-treated filter paper. The measurements were compared to those obtained using β ray method and were found consistent. This method can be particularly relevant to polluted atmospheres where PM is dominated by black carbon.