Automated PCR detection of BRAF mutations in colorectal adenocarcinoma: a diagnostic test accuracy study
Richard Colling1, Lai Mun Wang1, Elizabeth Soilleux2
1Department of Cellular Pathology, Oxford University Hospitals NHS Trust, Oxford, UK.
Background:
Testing for BRAF mutations in colorectal carcinoma (CRC) is important in the screening pathway for Lynch syndrome and is of prognostic value to guide management. This is a diagnostic accuracy study of the Idylla system, a novel and automated alternative PCR system.
Methods:
100 consecutive formalin-fixed, paraffin-embedded CRC resection cases were tested for BRAF mutations using the Idylla automated platform and compared with standard (Cobas) PCR.
Results:
The sensitivity of the Idylla BRAF test was 100% and the specificity was 96%. Only one discordant Idylla positive/standard PCR negative result occurred and on Droplet Digital PCR demonstrated a mutation not identified by traditional PCR in this case.
Conclusion:
This study has validated the Idylla system for BRAF testing in CRC and demonstrated a possibly greater sensitivity, in addition to cost effectiveness and shorter turnaround time, when compared with standard PCR.