Experiences of using a commercial dose management system (GE DoseWatch) for CT examinations
Ruth M Nicol1,2, Sarah C Wayte1, Andrew J Bridges1
11 Department of Clinical Physics and Bio-Engineering, Arden Cancer Centre, University Hospitals Coventry and Warwickshire NHS Trust, Coventry, UK.
Objective:
To assess the use of a commercial dose management system (GE DoseWatch; GEMS, Milwaukee, WI) for CT dose management for six common CT examinations.
Methods:
Data were acquired over several months using GE DoseWatch for six common CT examinations on three CT scanners. The dose length product (DLP) was taken as the dose indicator. The data were analysed using four different filtering methods: study description, the National Interim Clinical Imaging Procedure code, protocol name and a more detailed filtering method (the reference data set). The filtering methods were compared using an analysis of variance and multiple comparison technique. The different scanners were compared using the reference data set.
Results:
It was found that integrating DoseWatch with the radiology information system provided improved results compared with using the study description. Filtering by study description was found to be a poor indicator of the mean dose for all three scanners and consistently overestimated (p < 0.05) the head and thorax-abdo-pelvis mean DLP values, despite the large sample sizes. Filtering by the National Interim Clinical Imaging Procedure code or protocol name produced mean DLPs which were not statistically different from the reference data. The scanner intercomparison showed some significant differences between the scanners, usually due to different tube current modulation settings.
Conclusion:
The use of a commercial dose monitoring system provided fast and efficient filtering of substantial amounts of data. The filtering method affected the mean DLP value despite large sample sizes.
Advances In Knowledge:
Dose management systems are relatively new in the UK, and this article shares knowledge on the use of one system.
More Related Videos
10:33Expedited Radiation Biodosimetry by Automated Dicentric Chromosome Identification ADCI and Dose Estimation
Published on: September 4, 2017
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
Related Concept Videos
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System V: CT
Dose Size and Dosing Frequency: Determination Methods
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Dosage Regimen: Fixed Dose
Fixed-dose regimens can be used for various routes of administration, including intravenous (IV) injections and oral medications. For IV administration, a predetermined amount of the drug is...
Therapeutic Drug Monitoring: Overview and Classification